Description
The MB120 delivers functionality and precision that you can rely on, and is easy to operate and clean – you can test more samples in less time! Advanced features such as a temperature guide help to analyze optimum drying temperature, and methods and results storage allows statistical analysis of measurements. Three levels of user management ensure data safety - set up and lock temperature settings and method parameters with administrator access.
Features
Applications
- Percent Moisture Determination, Percent Moisture Regain Determination, Percent Solid Determination, Weighing
Display
- 4.3" TFT color touchscreen display
Operation
- Power cord (included)
Communication
- RS232 with GLP/GMP data output, USB Host and USB Device (included)
Construction
- Metal base, ABS top housing, halogen heat source, stainless steel pan support, stainless steel pan handler, in-use cover
Design Features
- Seven shut-off criteria (manual, timed, autofree %/s, autofree mg/s, A30, A60 and A90), four drying profiles (standard, fast, ramp and step), storage up to 100 methods and 1000 test results for all methods, user guide, 40°C to 230°C heating temperature range (1°C increments), 13 operational languages
Specifications
Maximum Capacity | 120 g |
Readability Moisture Content | 1 mg/0.01% |
Pan Size | 3.5 in (90 mm) |
Heater Type | Halogen Lamp |
Communication | USB host x 2; USB device; RS232 |
Dimensions | 7.09 in x 13.98 in x 8.31 in (180 mm x 355 mm x 211 mm) |
Display | Full-Color 4.3 in Touchscreen WQVGA |
Heating Profiles | Standard; Ramp; Step; Fast |
In-use cover | Included |
Moisture Range | 0 % – 100 % |
Net Weight | 11.53 lb (5.23 kg) |
Operating Range (Temp.) | 10 °C – 40 °C |
Power | 115 v, 60 Hz |
Sample Size | 0.5 g – 120 g |
Shut-Off Criteria | Manual; Timed; Automatic - Preset Weight Loss/Time; Custom - User Set Weight Loss/Time; Custom-User Set Moisture% Loss/Time |
Temperature Range | 40 °C – 230 °C |
Test Library | 100 |
Units of Measurement | Gram; %MC; %RG; %DC |
Working Environment | 50°F – 104°F, 85%RH, non-condensing (10°C – 40°C, 85%RH, non-condensing) |
Documents
MB120 Data Sheet
MB120 Manual